OPERATING MANUAL
BFS-MP FREEZING STAGES
FOR MICROTOMES
BFS-3MP, BFS-5MP, BFS-30MP
CONTENTS
SECTION PAGE
1.0 GENERAL DESCRIPTION 2
2.0 UNPACKING AND CHECKING 2
CONTENTS
3.0 SETTING UP THE SYSTEM 3
4.0 OPERATING INSTRUCTIONS 4
4.6 Setting Stage Temperature
4.7 Removing Specimen from Stage
4.8 Typical Stage Performance
5.0 TROUBLE SHOOTING 5
6.0 SPECIFICATIONS 7
7.0 MAINTENANCE, WARRANTY AND 8
SERVICE
Appendix 1: Operating Instructions for Physitemp 9
Pump and Tank Units
Appendix 2: Operating Instructions for ECV-1 10
Electrically Controlled Solenoid Valve
QX P:\CW shared\Operating Manuals\EB BFS-MP 6310 Rev4 prtlw 10/10
Vedere la pagina 0
1 2 3 4 5 6 ... 11 12

Sommario

Pagina 1 - FOR MICROTOMES

OPERATING MANUALBFS-MP FREEZING STAGESFOR MICROTOMESBFS-3MP, BFS-5MP, BFS-30MPCONTENTSSECTION PAGE1.0 GENERAL DESCRIPTION 22.0 UNPACKING AND CHECKING

Pagina 2

APPENDIX 1OPERATING INSTRUCTIONS FOR PHYSITEMP PUMP AND TANK UNIT, PTU-31. Unscrew cap. Fill reservoir with 5 gallons of distilled water.2. Connect ex

Pagina 3 - BFS/MP SERIES THERMAL STAGES

APPENDIX 2OPERATING INSTRUCTIONSECV-1 ELECTRICALLY OPERATED SOLENOID VALVEThe ECV-1 is supplied with an AC line cord and 12 VDC desktop power supply.1

Pagina 4 - CONNECTIONS

PHYSITEMP INSTRUMENTS INC154 Huron AvenueClifton NJ 07013 USATel: 973-779-5577Fax: 973-779-5954email: [email protected] MANUALBFS-MP SERIES F

Pagina 5 - FRONT PANEL CONTROLS

WARNING! DO NOT USE DRY ICE ON TOP OF THE STAGESeveral customers have reported using Dry Ice (CO2) to speed up thefreezing of tissue specimens mounted

Pagina 6 - 5.0 TROUBLESHOOTING

BFS/MP SERIES THERMAL STAGES1.0 GENERAL DESCRIPTION1.1 Physitemp manufactures three freezing stages for use with microtomes. All stages operate in the

Pagina 7

3.0 SETTING UP THE SYSTEMThis section gives instructions for set-up if the equipment is to be used with tap water: If a Pump and Tank Unit is to be us

Pagina 8 - 6.0 SPECIFICATIONS

4.0 OPERATING INSTRUCTIONS4.1 Ensure that the POWER Switch on the rear panel of the controller is in the OFF position4.2 Connect the AC power cord to

Pagina 9

4.7 REMOVING SPECIMEN FROM STAGEPress the Freeze/Thaw switch and hold in the QUICK THAW position for a few seconds. Allow thestage to warm up until sp

Pagina 10 - GENERAL INSTRUCTIONS

collar in an anti-clockwise direction. An ohmmeter will be required to make several readings on thestage connector. Once the connector has been remove

Pagina 11 - OPERATING INSTRUCTIONS

6.0 SPECIFICATIONS6.1 CONTROLLEROperating Range: BFS-3MP and 5MP: -35°C to ambientBFS-30MP: -40°C to ambientDigital Readout Resolution: 0.1°CAccurac

Pagina 12

7.0 MAINTENANCE, WARRANTY AND SERVICE7.1 MAINTENANCEThe stage needs no maintenance at all. It may be cleaned as necessary with a soft cloth, water or